A three-detector Positron Lifetime Setup Suited for measurements on irradiated steels

Luc Van Hoorebeke, Albert Fabry, Eric van Walle, José Van de Velde, Danny Segers, L.V. Dorikens-Vanpraet

    Research outputpeer-review

    Abstract

    An advanced positron lifetime setup is described , particularry suited to perform measurements on irradiated nuclear reactor pressure vessel steel. The equipment exists of three BaF2-detectors operated in triple coincidence mode to avoid the very disturbing Co-concentration in the life-time spectra. The use of BaF2 scintillators gives rise to an acceptable efficiency and a time resolution well below 300 ps.
    Original languageEnglish
    Pages (from-to)566-571
    Number of pages6
    JournalNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
    Volume371
    Issue number3
    DOIs
    StatePublished - 1996