A three-detector positron lifetime setup suited for measurements on irradiated steels

L. Van Hoorebeke, A. Fabry, E. Van Walle, J. Van De Velde, D. Segers, L. Dorikens-Vanpraet

    Research outputpeer-review

    Abstract

    An advanced positron lifetime setup is described, particularly suited to perform measurements on irradiated nuclear reactor pressure vessel steel. The equipment exists of three BaF2-detectors operated in triple coincidence mode to avoid the very disturbing 60Co-contribution in the lifetime spectra. The use of BaF2 scintillators gives rise to an acceptable efficiency and a time resolution well below 300 ps.

    Original languageEnglish
    Pages (from-to)566-571
    Number of pages6
    JournalNuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment