Calibration of solid state P-N detectors with alpha and heavy high-energy particles

J. Colard, J. Gal

    Research outputpeer-review

    Abstract

    The ionization defect for solid state P-N junction detectors was accurately measured for fission fragments. It was found to be 4.4 ± 1 MeV for the light fragments and 6.4 ± 1.9 MeV for the heavy fragments. A method to calculate the absorption of energy in the dead layer for these fragments was developed, so that the linearity of the pulse height versus energy curve, for heavy high-energy particles, could be conserved when calibrated with α particles.

    Original languageEnglish
    Pages (from-to)195-198
    Number of pages4
    JournalNuclear Instruments and Methods
    Volume16
    Issue numberC
    DOIs
    StatePublished - 1962