Efficiency calibration of BEGe and extended range detectors

    Research outputpeer-review

    Abstract

    Any approach to efficiency calibration of HPGe detectors sensitive to low-energy gamma-rays and X-rays has to deal with true coincidence summing. We developed a method based on the integration of the EFFTRAN code into the commercial Genie2K™ gamma-ray spectrometry software. EFFTRAN's role is to compute true coincidence summing corrections and to obtain satisfactory results for detectors sensitive to X-rays. EFFTRAN had to be upgraded to include a full simulation of the emission of X-rays. In this manner, efficiency calibrations were successfully set up for BeGe and XtRa detectors.
    Original languageEnglish
    Pages (from-to)356-360
    JournalApplied Radiation and Isotopes
    Volume87
    Issue number5
    DOIs
    StatePublished - May 2014

    Cite this