Efficiency calibration of Si(Li) detectors with X-ray reference sources at energies between 1 and 5 keV

B. Denecke, G. Grosse, U. Wätjen, W. Bambynek, C. Ballaux

Research outputpeer-review

Abstract

Fluorescence sources were prepared and their KX-ray-emission rates certified with an accuracy of better than 2% using a gas-flow proportional counter in a defined low-solid-angle setup. These sources were used for the efficiency calibration of a Si(Li) detector at the characteristic KX-ray energies of Al, Cl, Ca and Ti.

Original languageEnglish
Pages (from-to)474-479
Number of pages6
JournalNuclear Inst. and Methods in Physics Research, A
Volume286
Issue number3
DOIs
StatePublished - 15 Jan 1990
Externally publishedYes