Activation analysis provides a handy tool for elementary fluorine determination, with a relative error of 1-2 per cent. As fluorine standards, ammonium fluoride and p-fIuorobenzoic acid can be used. The values obtained for the fluorine contents show a good reproducibility with platinum as flux monitor. As a result of the analysing technique used, most interferences are negligible or can be eliminated and the radiation damage can be restricted to a minimum so that in most cases the compound is not lost for further use. The method is applied to determine the changes in fluorine content of teflon by pile irradiation.
Original language | English |
---|
Publisher | SCK CEN |
---|
Number of pages | 10 |
---|
State | Published - Jun 1962 |
---|
Externally published | Yes |
---|
Name | SCK•CEN Reports |
---|
Publisher | Studiecentrum voor Kernenergie |
---|
No. | BLG-182 |
---|