Abstract
The air kerma and absorbed dose to water standards for 60Co radiotherapy beams of 6 EURAMET NMIs were compared by circulating 2 transfer standards. The schedule for the measurements lasted from March 2013 to February 2015. The participants determined the calibration coefficients for the 2 transfer standards in both radiation quantities mentioned above. The comparison results are given in terms of air kerma and absorbed dose to water for both transfer standards as the ratio of the participants' calibration coefficients to the respective average values over all participants. Two participating NMIs, BEV and METAS, measured the transfer standards twice and three times, respectively. Hence, submeans were calculated and propagated as their respective repeated measurements are correlated.
The results are also linked to the BIPM.RI(I)-K1 and BIPM.RI(I)-K4 key comparisons through the linking laboratories BEV, LNE-LNHB and VSL for the air kerma and BEV, LNE-LNHB and METAS for the absorbed dose to water standards.
The results are also linked to the BIPM.RI(I)-K1 and BIPM.RI(I)-K4 key comparisons through the linking laboratories BEV, LNE-LNHB and VSL for the air kerma and BEV, LNE-LNHB and METAS for the absorbed dose to water standards.
Original language | English |
---|---|
Publisher | EURAMET - European Association of National Metrology Institutes |
Number of pages | 17 |
State | Published - 18 Mar 2020 |
Publication series
Name | EURAMET PROJECTS TC IR |
---|---|
No. | IR 1285 |