Experimental test of the "zero dead time" count-loss correction method on a digital gamma-ray spectrometer

Stefaan Pommé

    Research outputpeer-review

    Abstract

    A DSPECplus™ digital gamma-ray spectrometer is tested on its ability to compensate for count loss due to pulse pileup rejection and to assess the statistical uncertainty on the measured number of counts in the gamma-ray spectra. In the applied LEGe detector set-up, count loss is compensated correctly up to about 90% system dead time. This is done by the so-called "zero dead time" count loss correction method, involving the regular addition of virtual counts to the true counts according to a short-term ratio of real time to system live time. Besides performing add-n operations to the gamma-ray spectrum instead of add-1, the spectrometer also builds up a Σn2-spectrum. Experimental tests in this work confirm that the statistical counting uncertainty in small regions of interest in the gamma-ray spectrum is well predicted from the sum of counts in the corresponding region of interest in the Σn2-spectrum. A correction formula is suggested to extend the validity of this method to large parts of the spectrum.

    Original languageEnglish
    Pages (from-to)245-252
    Number of pages8
    JournalNuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
    Volume474
    Issue number3
    DOIs
    StatePublished - 11 Dec 2001

    ASJC Scopus subject areas

    • Nuclear and High Energy Physics
    • Instrumentation