Experimental validation of corrections factors for γ–γ and γ–X coincidence summing of 133Ba, 152Eu, and 125Sb in volume sources

S. Jonsson, Johan Kastlander, Tim Vidmar, H. Ramebäck

    Research outputpeer-review

    Abstract

    True coincidence summing correction factors for 133Ba, 152Eu and 125Sb were determined experimentally for a small volume source and compared with correction factors obtained with three softwares (EFFTRAN-X, GESPECOR and VGSL). The radionuclides investigated have a relatively challenging decay scheme and their spectra are known to suffer from losses due to summation (γ–γ, γ–X and X–X) when measured at close distances on a HPGe detector sensitive to low energy photons. This study shows that the softwares were in good agreement with each other and the experimental data and the calculated activity was consistent with the activity in the volume source.
    Original languageEnglish
    Pages (from-to)465–472
    Number of pages8
    JournalJournal of Radioanalytical and Nuclear Chemistry
    Volume323
    DOIs
    StatePublished - 1 Jan 2020