Abstract
In the context of the decrease in the eye lens dose limit for occupational exposure to 20 mSv per year stated by the recent revision of the European Basic Safety Standards Directive 2013/59/EURATOM, the European Radiation Dosimetry Group (EURADOS) has organised in 2014, for the first time, an intercomparison exercise for eye lens dosemeters. The main objective was to assess the capabilities of the passive eye lens dosemeters currently in use in Europe for occupational monitoring in medical fields. A total of 20 European individual monitoring services from 15 different countries have participated. The dosemeters provided by the participants were all composed of thermoluminescent detectors, of various types and designs. The irradiations were carried out with several photon fields chosen to cover the energy and angle ranges encountered in medical workplace. Participants were asked to report the doses in terms of Hp(3) using their routine protocol. The results provided by each participant were compared with the reference delivered doses. All the results were anonymously analysed. Results are globally satisfactory since, among the 20 participants, 17 were able to provide 90 % of their response in accordance with the ISO 14146 standard requirements.
Original language | English |
---|---|
Pages (from-to) | 21-26 |
Number of pages | 6 |
Journal | Radiation Protection Dosimtery |
Volume | 170 |
Issue number | 1-4 |
DOIs | |
State | Published - 1 Sep 2016 |
Event | International Conference on Individual Monitoring of Ionising Radiation - Eurados, Bruges Duration: 20 Apr 2015 → 24 Apr 2015 |