Abstract
Intercomparison exercises (IE) are a way for laboratories to test their analytical performance. In case of difficult-to-measure (DTM) radionuclide analysis, participation in IEs are important due to the lack of reference materials. This paper reports the results from an IE focusing on non-volatile DTM analysis in homogenised high activity steel. The IE was carried out according to the ISO 13528 standard, in which the performances are assessed using z score. The z scores were generally acceptable. The experimental results were compared with activation calculation results showing wide range of compatibility.
| Original language | English |
|---|---|
| Pages (from-to) | 351-374 |
| Number of pages | 24 |
| Journal | Journal of Radioanalytical and Nuclear Chemistry |
| Volume | 335 |
| DOIs | |
| State | Published - 16 Jan 2026 |
ASJC Scopus subject areas
- Analytical Chemistry
- Nuclear Energy and Engineering
- Radiology Nuclear Medicine and imaging
- Pollution
- Spectroscopy
- Public Health, Environmental and Occupational Health
- Health, Toxicology and Mutagenesis
Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver