Abstract
The γ-ray emission rate of 109Cd can be measured by means of spectrometry with a NaI(Tl) well-type detector. The corrections from photoelectron escape, from Compton effect near the interface crystal-liner, from bremsstrahlung, produced during slowing-down in NaI(Tl) and by the conversion electrons, and from other minor effects have been calculated.
Original language | English |
---|---|
Pages (from-to) | 923-928 |
Number of pages | 6 |
Journal | International Journal of Radiation Applications and Instrumentation. Part A. Applied Radiation and Isotopes |
Volume | 37 |
Issue number | 9 |
DOIs | |
State | Published - 1986 |
Externally published | Yes |
ASJC Scopus subject areas
- Radiation