Abstract
Within the EC-funded CHANCE project several non-destructive techniques are being considered for the assay of waste bearing drums. Such techniques include calorimetry, gamma-ray spectrometry and neutron coincidence counting. The aim is to quantify uncertainties on the inventory of radionuclides, and how these are potentially reduced by combining the signatures from different techniques in the data analysis.
In this framework, neutron coincidence measurements were carried out with two slab counters based on 3He detectors coupled to shift register electronics. Such a system consists of two identical slabs with 6 detectors each, and is transportable, rather compact and flexible in terms of sizes and geometries that can be measured. With this system three 200 L drums containing certified reference nuclear material and different filling materials were measured. The certified nuclear material was in the form of 21 pellets of mixed oxide of U and Pu with a total mass of about 10.5 g; in addition, a single pellet of about 10.05 g was also available. The pellets could be placed in predefined positions within the drum in a reproducible way. The geometry and composition of the three drums was well characterized and consisted of Ethafoam, a mixture of Ethafoam, stainless steel and PVC, and mortar with an inner core of extruded polystyrene. The measurement setup was arranged such that the drum was placed between the two slab counters. The positions of the slab counters relative to the drum were accurately measured before each measurement, and a dedicated system was used to minimize the uncertainty on the detector positioning.
The measurement data were first analysed by applying the point model of Hage and the mass of nuclear material in the drum was determined from the rate of totals and reals and the radionuclide composition. Due to the fact that not all the point model conditions were met, we found that the point model overestimates the mass up to about 50%. In addition, a Monte Carlo model of the measurement geometry was developed using the MCNP code. The model was used to determine a calibration factor between the reals rate and the mass of the sample. Measurements with a calibrated 252Cf source were used to verify the model. With a Monte Carlo based approach the mass of the mixed oxide pellets is within a few percent from the nominal values, except for strongly asymmetrical configurations where the deviation is up to about 20%. The results reveal the importance of an accurate background correction and of accounting for surrounding materials of the building such as walls, floor and ceiling in the Monte Carlo model.
Original language | English |
---|---|
Article number | 07001 |
Pages (from-to) | 1-7 |
Number of pages | 7 |
Journal | EPJ Web of Conferences |
Volume | 253 |
DOIs | |
State | Published - 19 Nov 2021 |
Event | 2021 - ANIMMA - Advancements in Nuclear Instrumentation Measurement Methods and their Applications - CLARION Congress Hotel Prague, Prague Duration: 21 Jun 2021 → 25 Jun 2021 Conference number: 7 https://indico.utef.cvut.cz/event/23/ |