Abstract
Fibre Bragg Grating (FBG) sensors are expected to provide valuable data in extreme radiation environments associated with nuclear research reactors. However, when the fast neutron fluence reaches 1018 to 1019 n/cm², the radiation induced changes in the material density and refractive index may drastically bias the measurements. The present study evaluates the radiation effect on the FBG performances by comparing their properties before and after their exposure to fast neutron fluences exceeding 1019 n/cm² (E > 1 MeV). We studied responses of FBGs manufactured by three different laboratories in the same single-mode optical fibre but using different inscription conditions. The Bragg wavelength and the reflectivity were measured before and after irradiation thanks to a dedicated mounting. For nearly all FBGs, the Bragg peak remains visible after the irradiation while the radiation-induced Bragg wavelength shifts (RI-BWS) vary from a few pm (equivalent temperature error <1°C) to nearly 1 nm (~100°C error) depending of the FBG inscription conditions. Such high RI-BWSs can be explained by the huge refractive-index variation and compaction observed for bare fibre samples through other experimental techniques. Our results show that by using specific hardening techniques the FBG-based temperature measurements in a nuclear research reactor experiment may become feasible.
Original language | English |
---|---|
Article number | 102 |
Pages (from-to) | 1-8 |
Number of pages | 8 |
Journal | IEEE transactions on nuclear Science |
Volume | 65 |
Issue number | 9 |
DOIs | |
State | Published - 21 Feb 2018 |