Summary of personal neutron dosemeter results obtained within the EVIDOS project

M. Luszik-Bhadra, T. Bolognese-Milsztajn, M. Boschung, Michèle Coeck, G. Curzio, D. Derdau, F. d'Errico, A. Fiechtner, J.E. Kyllönen, V. Lacoste, B. Lievens, L. Lindborg, A. Loverfors Daun, M. Reginatto, H. Schuhmacher, R. Tanner, Filip Vanhavere

    Research outputpeer-review

    Abstract

    Within the Ec project EVIDOS ('Evaluation of Individual Dosimetry in Mixed Neutron and Photon Radiation fields') different types of active neutron personal dosemeters (and some passive ones) were tested in workplace fields at nuclear installations in Europe. The results of the measurements which have been performed up to now are summarised and compared to our currently best estimates of the personal dose equivalent Hp(10). Under- and over-readings by more than a factor of two for the same dosemeter in different workplace fields indicate that in most cases the use of field-specific correction factors is required.
    Original languageEnglish
    Pages (from-to)293-299
    JournalRadiation protection dosimetry
    Volume125
    Issue number1-4