The life span of the BD-PND bubble detector

F. Vanhavere, M. Loos, H. Thierens

    Research outputpeer-review

    Abstract

    BD-PND bubble detectors from Bubble Technology industries (BTI) were used to conduct a study of the life span of these detectors. The manufacturer guarantees an optimum detector performance for three months after receipt. Neuertheless, it is important to know the evolution of their characteristics with time, also after those three months. On a standard set-up with a 252Cf source the bubble detectors were irradiated until they reached the end of their life span. During this period, the evolution in sensitivity was monitored. The temperature compensating system seems to be the limiting factor with time for the use of the BTI bubble detectors. The change in temperature dependence with age was determined. The same parameters were also checked with several batches of detectors that were used in practice.

    Original languageEnglish
    Pages (from-to)27-30
    Number of pages4
    JournalRadiation protection dosimetry
    Volume85
    Issue number1-4
    DOIs
    StatePublished - 1999