Abstract
The fabrication and quality control of Ni-3 wt.% U (93% 235U) alloy rods (diameter: 8 mm, total length: 8 m) used as a reference fission product source (RFPS) are described. This RFPs is needed for the calibration of the fuel element rupture detector which is used for the continuous monitoring of the delayed neutron activity in the primary sodium of the Super Phenix fast neutron reactor. Special care was taken to obtain a sufficient alloy purity, a bright surface finish and a homogeneous 235U content on the surface of the rods. Finally the usefulness of the RFPS in Super Phenix has been demonstrated.
Original language | English |
---|---|
Pages (from-to) | 35-39 |
Number of pages | 5 |
Journal | Nuclear Inst. and Methods in Physics Research, A |
Volume | 257 |
Issue number | 1 |
DOIs | |
State | Published - 1 Jun 1987 |
ASJC Scopus subject areas
- Nuclear and High Energy Physics
- Instrumentation